Introduction
Used for product simulated hydraulic environment testing. It uses compressed air as the power source and a gas-liquid booster pump as the pressure source. The input water pressure is 80 times proportional to the driving air source pressure. By adjusting the driving air source pressure, the corresponding boosted water pressure can be obtained. When the driving air source pressure is balanced with the boosted liquid pressure, the booster pump stops charging and the input water pressure stabilizes at the preset pressure. Therefore, it has the characteristics of explosion-proof, adjustable output pressure, small size, light weight, simple operation, reliable performance and wide range of use.
Features
1. The booster unit is installed independently, and the connection is safe and reliable. All instruments that need to be observed and operated are centrally installed on the panel. The operator can monitor the operating status of the equipment through external observation, enabling unattended work.
2. The system itself has a built-in high-pressure one-way valve; ensuring the safe and reliable operation of each boosting unit without affecting each other;
3. By adjusting the compressed air pressure, the output hydraulic pressure can be continuously adjusted within the pressure output range;
4. Set the upper and lower limits of supercharging through the pressure controller;
5. Low noise, small equipment size, low installation cost and low maintenance cost.
Pressurized medium | water, alcohol |
Power source | 0.4-0.8Mpa clean compressed air |
Output pressure calculation formula | 80 times driving air pressure |
Maximum test pressure | 35MPa |
Maximum flow | 2.4L/min |
Maximum air consumption | 1.2M3/[email protected] |
Compressed air interface | Φ12 quick-plug connector |
Liquid high pressure outlet | NP1/4 internal thread |
Hyperbaric chamber size | inner diameter 200mm, height 300mm |